Report Number
8030965-2014-00806
Event Type
Malfunction
Date Received
September 19, 2014
Report Date
February 15, 2013
Manufacturer
SYNTHES GMBH
Product Code
HWE
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO SYNTHES OUS SERVICE AND REPAIR FILES LEGACY REVIEW/REMEDIATION PROTOCOL-COMPLAINT HANDLING AND MDR REPORTING. DEVICE LISTED IN THIS REPORT IS USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. DURING PRE-REPAIR ASSESSMENT PERFORMED BY A TECHNICIAN, FUNCTIONAL TESTING WAS PERFORMED AND REVEALED THE ADAPTOR HAS THE CONNECTION PART BROKEN DUE TO THE USE OF EXCESSIVE FORCE AND INCORRECT MOVEMENTS IN THE PROCEDURE, CREATING STRESS IN A CERTAIN POINT AND POSTERIOR BREAK. DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER ON (B)(4) 2013. SECOND SERIAL NUMBER IS (B)(4). PLACEHOLDER.

Description of Event or Problem · 1

DEVICE WAS RETURNED FOR AN UNKNOWN REASON. DURING THE SERVICE TECHNICIAN FOUND THAT THE CONNECTION PART OF THE MANDREL WAS BROKEN. THIS IS REPORT 1 OF 1 FOR COMPLAINT#(B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
584326 DRILL CHUCK KEYLESS F/COMPACT AIR DRIVE+ HWE SYNTHES GMBH 10640

Patients

Seq Age Sex Outcome Treatment
1