FDA Adverse Event Malfunction Summary report: N

4FR PICC BASIC TRAY

MDR report key: 4102727 · Received August 28, 2014

Report

Report Number
3006260740-2014-00437
Event Type
Malfunction
Date Received
August 28, 2014
Date of Event
February 26, 2014
Report Date
February 26, 2014
Manufacturer
BARD ACCESS SYSTEMS INC.
Product Code
DQY
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE COMPLIANT OF INABILITY TO ASPIRATE WAS CONFIRMED AND FOUND TO BE USE-RELATED. THE RETURNED PRODUCT WAS ONE 4FR S/L GROSHONG PICC. THE STYLET WAS STILL INLAID WITHIN THE CATHETER AND THE SAMPLE WAS FREE OF RESIDUE. AN ATTEMPT TO INFUSE AND ASPIRATE WATER THROUGH THE CATHETER REVEALED WATER ASPIRATION WITH AN AIR LEAK, AND A SPRAYING LEAK WAS FOUND AT THE 46CM DEPTH MARKING. MICROSCOPIC EXAMINATION OF THE LEAK SITE REVEALED A SMALL IRREGULAR SPLIT. WHEN THE STYLET WAS REMOVED AND THE CATHETER SPLIT LONGITUDINALLY, THE INTERIOR OF THE CATHETER EXHIBITED INTERMITTENT IMPRESSIONS TYPICAL OF STYLET DAMAGE. TACTILE EVALUATION OF THE CATHETER WAS UNREMARKABLE. THE DAMAGE FOUND AT THE LEAK SITE WAS TYPICAL OF STYLET DAMAGE BY THE USER.

Description of Event or Problem · 1

IT WAS REPORTED THAT NO BACK FLOW IN PICCLINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
526744 4FR PICC BASIC TRAY DQY BARD ACCESS SYSTEMS INC. REW11275

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention