FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 4102719 · Received September 19, 2014

Report

Report Number
2531779-2014-26840
Event Type
Malfunction
Date Received
September 19, 2014
Report Date
September 9, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

FOLLOW-UP #1 DATE OF SUBMISSION (B)(6) 2015-PRODUCT ANALYSIS: THE DEVICE WAS RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(6) 2015 WITH THE FOLLOWING FINDINGS: INVESTIGATION REVEALED THE DISPLAY SCREEN WAS DIM AND DISCOLORED. UNRELATED TO THE COMPLAINT, A BATTERY COMPARTMENT CRACK WAS OBSERVED.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (DIM/FADING/COLOR SPECTRUM) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
584932 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 77 YR