ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-26839
- Event Type
- Malfunction
- Date Received
- September 19, 2014
- Report Date
- September 8, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- PATIENT
Narratives
FOLLOW-UP #1: DATE OF SUBMISSION 11/20/2014. DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: THERE WERE PUMP REBOOTS OBSERVED IN THE BLACK BOX HISTORY. THE PUMP POWERED ON WITH THE RETURNED BATTERY CAP. THE CAP WAS ABLE TO FULLY TIGHTEN TO THE PUMP AND MEASURED WITHIN SPECIFICATIONS. A POWER LOSS WAS NOT DUPLICATED. THE PUMP WAS EXERCISED FOR 24 HOURS WITH NO REBOOTS, LOSS OF POWER OR CALL SERVICE ALARMS DUPLICATED. THE PUMP CASE WAS REMOVED AND THERE WAS NO EVIDENCE OF MOISTURE OR LOOSE COMPONENTS INSIDE THE PUMP. UNRELATED TO THE ORIGINAL COMPLAINT, THE BATTERY COMPARTMENT WAS CRACKED.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (INTERMITTENT POWER) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 584171 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 15 YR |