PLUMSET CLAVE PORT CLAVE SITE 103IN NDHP
Report
- Report Number
- 9615050-2014-05028
- Event Type
- Malfunction
- Date Received
- August 28, 2014
- Date of Event
- August 19, 2014
- Report Date
- August 18, 2014
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K982159
- Removal / Correction Number
- NA
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- OTHER
Narratives
THE DEVICE IS EXPECTED TO BE RETURNED FOR INVESTIGATION. IT HAS NOT YET BEEN RECEIVED. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
THE CUSTOMER CONTACT REPORTED BREAKAGE OF THE DISTAL TIP OF THE OPTION-LOK MALE ADAPTER. THE PLUMSET WAS BEING USED TO DELIVER AN UNSPECIFIED MEDICATION AT AN UNSPECIFIED RATE. AFTER AN UNSPECIFIED LENGTH OF TIME, IT WAS REPORTED THAT THE DISTAL TIP OF THE OPTION-LOK MALE ADAPTER BROKE OFF AND A PIECE REMAINED LODGED INSIDE THE HUB OF THE PATIENT'S IV ACCESS DEVICE. NO SPECIFIC DETAILS WERE PROVIDED. THE PLUMSET WAS REPLACED AND THE THERAPY WAS RESUMED. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EFFECTS AND NO REPORTED DELAY IN THERAPY CRITICAL TO THIS PATIENT. NO MEDICAL INTERVENTIONS WERE REQUIRED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 523723 | PLUMSET CLAVE PORT CLAVE SITE 103IN NDHP | 80FRN | FRN | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNSPECIFIED IV ACCESS DEVICE, MFR UNK |