Narratives
Additional Manufacturer Narrative · 1
A FIELD SERVICE ENGINEER PERFORMED TESTING AND INVESTIGATION AT THE USER FACILITY, DURING TESTING A REVIEW OF THE DEVICE HISTORY INDICATED A S233 (OVERTEMPERATURE-PSC) MALFUNCTION ALARM CODE. THE PROBABLE CAUSE OF THE S233 MALFUNCTION ALARM CODE WAS A BROKEN FAN. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
Description of Event or Problem · 1
THE CUSTOMER CONTACT REPORTED THAT THE DEVICE ALARMED WITH A S233 (OVER TEMPERATURE- PSC) MALFUNCTION ALARM CODE. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS AND NO REPORTED DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. DURING TESTING AT THE USER FACILITY, A S233 MALFUNCTION ALARM CODE WAS NOTED. NO ADDITIONAL INFORMATION WAS PROVIDED.