FDA Adverse Event
Malfunction
Summary report: N
SYMBIQ DCHANNEL 3.01
MDR report key: 4102382
·
Received August 27, 2014
Report
- Report Number
- 9615050-2014-05000
- Event Type
- Malfunction
- Date Received
- August 27, 2014
- Date of Event
- August 11, 2014
- Report Date
- August 11, 2014
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K110901
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE IS EXPECTED TO BE RETURNED FOR INVESTIGATION. IT HAS NOT YET BEEN RECEIVED. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
Description of Event or Problem · 1
THE CUSTOMER CONTACT REPORTED THAT DURING PREVENTATIVE MAINTENANCE TESTING AT THE USER FACILITY, THE DEVICE DID NOT PASS THE DELIVERY ACCURACY TEST. IT WAS REPORTED THE DEVICE DELIVERED LESS THAN EXPECTED. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS AND NO REPORTED DELAYS OF CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. NO ADDITIONAL INFO WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 522606 | SYMBIQ DCHANNEL 3.01 | 80FRN | FRN | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |