FDA Adverse Event Malfunction Summary report: N

HUT EXT DR FINAL ASSY - STANDARD

MDR report key: 4102327 · Received August 27, 2014

Report

Report Number
1518293-2014-00100
Event Type
Malfunction
Date Received
August 27, 2014
Date of Event
July 30, 2014
Report Date
July 30, 2014
Manufacturer
LIEBEL FLARSHEIM
Product Code
IXR
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FIELD SERVICE ENGINEER (FSE) INVESTIGATED AND CONFIRMED THE TABLE MOTIONS COMPLAINT, TROUBLESHOT AND REPLACED THE 24V POWER SUPPLY. AFTER REPLACING THE POWER SUPPLY, FSE FOUND THAT ONLY HALF OF TABLE MOTIONS WERE OPERABLE. FURTHER TROUBLESHOOTING SHOWED FAILURE OF THE HYDRAULIC PUMP MOTOR CONTROLLER. FSE REPLACED THE HYDRAULIC PUMP MOTOR CONTROL ASSEMBLY AND THE TABLE OPERATED NORMALLY. THE UNIT WAS CHECKED FOR PROPER OPERATION PER SERVICE MANUAL AND RETURNED TO FULL SERVICE.

Description of Event or Problem · 1

CUSTOMER REPORTS VIA PHONE THAT DURING AN KIDNEY STONE REMOVAL PROCEDURE, PATIENT GENDER AND AGE NOT REPORTED BY THE CUSTOMER, THE TABLE MOVEMENTS FAILED. STAFF MOVED THE PATIENT TO ANOTHER ROOM, BUT THEY MOVED THE PATIENT BACK TO THE UROLOGY ROOM, AND USED THE TABLE "OVER-RIDE" BUTTON TO COMPLETE THE PROCEDURE. NO REPORTED INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
522663 HUT EXT DR FINAL ASSY - STANDARD IXR LIEBEL FLARSHEIM DR FINAL A HUT EXT NA

Patients

Seq Age Sex Outcome Treatment
1 UNK