FDA Adverse Event
Malfunction
Summary report: N
TRILOGY 100
MDR report key: 4102308
·
Received August 27, 2014
Report
- Report Number
- 2518422-2014-01473
- Event Type
- Malfunction
- Date Received
- August 27, 2014
- Date of Event
- July 31, 2014
- Report Date
- July 31, 2014
- Manufacturer
- RESPIRONICS INC.,
- Product Code
- CBK
- PMA / PMN Number
- K083526
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
THE VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR EVALUATION AND THE CUSTOMER'S COMPLAINT WAS CONFIRMED. THE DEVICE'S INTERFACE BOARD WAS REPLACED TO ADDRESS THE ISSUE.
Description of Event or Problem · 1
THE MANUFACTURER RECEIVED INFORMATION FROM A THIRD PARTY SERVICE CENTER ALLEGING A VENTILATOR'S REMOTE ALARM CONNECTOR WAS NOT WORKING. THE DEVICE WAS NOT IN PATIENT USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 522689 | TRILOGY 100 | VENTILATOR, CONTINUOUS, FACILITY USE | CBK | RESPIRONICS INC., | 1054260 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |