FDA Adverse Event Injury Summary report: N

XIENCE XPEDITION EVEROLIMUS ELUTING CORONARY STENT SYSTEM

MDR report key: 4102307 · Received September 19, 2014

Report

Report Number
2024168-2014-06048
Event Type
Injury
Date Received
September 19, 2014
Date of Event
August 21, 2014
Report Date
August 27, 2014
Manufacturer
AV-TEMECULA-CT
Product Code
NIQ
PMA / PMN Number
P110019
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR ANALYSIS. THE SHAFT SEPARATION WAS CONFIRMED. THE FAILURE TO ADVANCE COULD NOT BE REPLICATED IN A TESTING ENVIRONMENT AS IT WAS BASED ON OPERATIONAL CIRCUMSTANCES. BASED ON A VISUAL AND DIMENSIONAL INSPECTION OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. A QUERY/REVIEW OF THE COMPLAINT DATABASE REVEALED NO OTHER SIMILAR INCIDENTS REPORTED FROM THIS LOT. BASED ON THE REVIEWED INFORMATION, NO PRODUCT DEFICIENCY WAS IDENTIFIED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A HEAVILY CALCIFIED LESION IN THE HEAVILY TORTUOUS PROXIMAL RIGHT CORONARY ARTERY (RCA). THE 3.25X15MM XIENCE XPEDITION STENT DELIVERY SYSTEM (SDS) FAILED TO CROSS THE LESION REPORTEDLY DUE TO INTERACTIONS WITH THE PATIENT'S ANATOMY AND WHILE TORQUING, THE SDS MID-SHAFT SEPARATED. THE PROXIMAL PORTION OF THE SEPARATED SDS WAS WITHDRAWN WITHOUT ISSUE, AND A SNARE DEVICE WAS USED TO RETRIEVE THE DISTAL PORTION OF THE SEPARATED SDS. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. A 3.25X12MM XIENCE XPEDITION SDS WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. THERE WAS NO ADDITIONAL INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
584102 XIENCE XPEDITION EVEROLIMUS ELUTING CORONARY STENT SYSTEM DRUG ELUTING CORONARY STENT SYSTEM NIQ AV-TEMECULA-CT 4062061

Patients

Seq Age Sex Outcome Treatment
1 84 YR Required Intervention