FDA Adverse Event
Malfunction
Summary report: N
TRILOGY 100
MDR report key: 4102299
·
Received August 27, 2014
Report
- Report Number
- 2518422-2014-01472
- Event Type
- Malfunction
- Date Received
- August 27, 2014
- Date of Event
- August 19, 2014
- Report Date
- August 19, 2014
- Manufacturer
- RESPIRONICS INC.,
- Product Code
- CBK
- PMA / PMN Number
- K083526
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, A FAILURE OF THE DEVICE TO POWER ON AND CHARGE IT'S INTERNAL BATTERY WAS OBSERVED. THE DEVICE'S SYSTEM BOARD AND POWER MANAGEMENT BOARD WERE REPLACED TO ADDRESS THE ISSUES. THE DEVICE FAILED A STEP DURING TESTING. THE DEVICE'S ACTIVE EXHALATION CONTROL MODULE WAS REPLACED TO ADDRESS THE ISSUE.
Description of Event or Problem · 1
A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR SERVICE. THERE WAS NO HARM OR INJURY REPORTED.