FDA Adverse Event Malfunction Summary report: N

TRILOGY 100

MDR report key: 4102299 · Received August 27, 2014

Report

Report Number
2518422-2014-01472
Event Type
Malfunction
Date Received
August 27, 2014
Date of Event
August 19, 2014
Report Date
August 19, 2014
Manufacturer
RESPIRONICS INC.,
Product Code
CBK
PMA / PMN Number
K083526
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, A FAILURE OF THE DEVICE TO POWER ON AND CHARGE IT'S INTERNAL BATTERY WAS OBSERVED. THE DEVICE'S SYSTEM BOARD AND POWER MANAGEMENT BOARD WERE REPLACED TO ADDRESS THE ISSUES. THE DEVICE FAILED A STEP DURING TESTING. THE DEVICE'S ACTIVE EXHALATION CONTROL MODULE WAS REPLACED TO ADDRESS THE ISSUE.

Description of Event or Problem · 1

A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR SERVICE. THERE WAS NO HARM OR INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
522687 TRILOGY 100 VENTILATOR, CONTINUOUS, FACILITY USE CBK RESPIRONICS INC., 1054260

Patients

Seq Age Sex Outcome Treatment
1