FDA Adverse Event Malfunction Summary report: N

STAPLE, IMPLANTABLE

MDR report key: 4102136 · Received September 19, 2014

Report

Report Number
3005075853-2014-06530
Event Type
Malfunction
Date Received
September 19, 2014
Date of Event
August 14, 2014
Report Date
August 14, 2014
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K110385
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). SHOULD THE INFORMATION BE PROVIDED LATER, A SUPPLEMENTAL MEDWATCH WILL BE SENT. THE ANALYSIS FOUND THAT ONE PCE60A DEVICE WAS RECEIVED FOR ANALYSIS WITHOUT CARTRIDGE RELOAD. FURTHERMORE, THE DEVICE WAS FOUND TO HAVE THE CLAMPING MECHANISM DAMAGED. IN ORDER TO EVALUATE THE CONDITION OF THE INTERNAL COMPONENTS, THE DEVICE WAS DISASSEMBLED. UPON DISASSEMBLING, THE CLAMP ARM RELEASE AND RELEASE BUTTON WERE NOTED WORN AND DAMAGED IN THE AREA WHERE BOTH INTERACT. THIS DAMAGE SUGGESTS THE DEVICE WAS FORCED TO OPEN WITH THE KNIFE NOT IN THE HOME POSITION BY PULLING THE CLOSING TRIGGER TO HOME. TO OPEN THE JAWS, SQUEEZE THE CLOSING TRIGGER, AND THEN SIMULTANEOUSLY PRESS THE ANVIL RELEASE SWITCH ON EITHER SIDE OF THE INSTRUMENT. WHILE PRESSURE IS STILL ON THE ANVIL RELEASE SWITCH, SLOWLY RELEASE THE CLOSING TRIGGER. NO FURTHER FUNCTIONAL TEST COULD BE PERFORMED DUE TO THE CONDITION OF THE DEVICE. HOWEVER A DRY FIRE WAS PERFORMED AND THE FIRING MECHANISM WORKED AS INTENDED. IT SHOULD BE NOTED THAT AS PART OF OUR QUALITY PROCESS, EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED DURING MANUFACTURING TO ENSURE THAT THE DEVICE MEETS THE REQUIRED SPECIFICATIONS PRIOR TO SHIPMENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, AFTER SEVERAL FIRINGS, THE SURGEON COULD NOT CLOSE THE JAW OF THE DEVICE. IT IS UNKNOWN HOW THE PROCEDURE WAS COMPLETED. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT REPORTED.