NURSE CALL CABLE
Report
- Report Number
- 2020362-2014-00306
- Event Type
- Malfunction
- Date Received
- September 15, 2014
- Report Date
- August 13, 2014
- Manufacturer
- J. T. POSEY CO.
- Product Code
- KMI
- PMA / PMN Number
- EXEMPT
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- UNKNOWN
Narratives
RESULTS: EVAL FOUND THAT DURING TESTING, THE NURSE CALL CABLE WITH THE ALARM, WHEN THE NURSE CALL CABLE IS WIGGLED, THE NURSE CALL LIGHT TOGGLES ON AND OFF AT THE TEST FIXTURE. WHEN THE NURSE CALL CABLE IS USED WITH A KNOWN GOOD ALARM, THE NURSE CALL LED TURNS ON AT THE NURSE CALL TEST FIXTURE WHEN THERE IS NO WEIGHT ON THE SENSOR (AS EXPECTED) AND THE NURSE CALL LED TURNS OFF WHEN WEIGHT IS ON THE SENSOR (AS EXPECTED). HOWEVER, IF THE CABLE IS WIGGLED, THE NURSE CALL LED LIGHT TOGGLES ON AND OFF. THE NURSE CALL CABLE WAS SCUFF MARKS ON IT AND ONE OF THE JACKS IS LOOSE. WHEN THE NURSE CALL CABLE WAS OPENED, IT WAS FOUND TO HAVE A BROKEN GROUND WIRE. (B)(4).
THE CUSTOMER REPORTED THE NURSE'S STATION SOUNDS INCONSISTENTLY WHEN THE ALARM IS IN USE WITH THE NURSE CALL CABLE, DISCOVERED WHILE IN USE BUT CUSTOMER COULD NOT PROVIDE THE DATE OF THE EVENT. NO PT INCIDENT OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 569317 | NURSE CALL CABLE | KMI | J. T. POSEY CO. | 8282 | N1147 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | SITTER ELITE ALARM: MODEL 8345, SN # (B)(4) |