FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 4101890 · Received August 25, 2014

Report

Report Number
8020893-2014-02117
Event Type
Malfunction
Date Received
August 25, 2014
Date of Event
August 11, 2014
Report Date
August 16, 2014
Manufacturer
COVIDIEN
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) EVALUATED THE DEVICE, AND VERIFIED THE MALFUNCTION. THE CSE REPLACED THE KEYBOARD TO RESOLVE THE ISSUE. THE UNIT PASSED ALL TESTS AND CALIBRATIONS, AND IT WAS OBSERVED TO BE OPERATING WITHIN MANUFACTURING SPECIFICATIONS. (B)(4).

Additional Manufacturer Narrative · 1

COVIDIEN RECEIVED A KEYBOARD PRINTED CIRCUIT BOARD (PCB), WHICH WAS REPORTED AS HAVING A NON-OPERATIONAL RESET KEY. A VISUAL INSPECTION OF THE RETURNED KEYBOARD SHOWED NO ANOMALIES. FUNCTIONALITY WAS PERFORMED BY USING A FAILURE INVESTIGATION TEST VENTILATOR FOR ANALYSIS. TEST VENTILATOR WAS VERIFIED BY SUCCESSFULLY RUNNING ALL TESTS AND CALIBRATION AS DETAILED IN THE 840 VENTILATOR SERVICE MANUAL. THE KEYBOARD WAS ATTACHED TO THE UNIT, POWERED UP NORMALLY, RUN SUCCESSFULLY, AND NO ERRORS WERE RECORDED IN THE DIAGNOSTIC LOGS. THE UNIT WAS SET INTO NORMAL VENTILATION MODE, AND RAN SUCCESSFULLY FOR OVER 24 HOURS. A REVIEW OF THE DIAGNOSTIC LOGS REVEALED NO ERRORS AND NO ALARMS WERE OBSERVED DURING THE RUN IN VENTILATION MODE PERIOD. PREVIOUS INVESTIGATIONS INDICATED THAT THE PROBABLE ROOT CAUSE OF THE REPORTED FAILURE COULD BE CAUSED BY CONTAMINATION ON THE PCB. THE RETURNED KEYBOARD WAS FOUND FUNCTIONING AS PER SPECIFICATIONS, AND THE REPORTED MALFUNCTION COULD NOT BE DUPLICATED.

Description of Event or Problem · 1

IT WAS REPORTED BY A CUSTOMER IN THE UNITED KINGDOM THAT THE 840 VENTILATOR RESET KEY WAS INOPERATIVE. THERE WAS NO PT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
511166 840 VENTILATOR CONTINUOUS VENTILATOR CBK COVIDIEN 840

Patients

Seq Age Sex Outcome Treatment
1