FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 4101854 · Received September 19, 2014

Report

Report Number
1416980-2014-32220
Event Type
Malfunction
Date Received
September 19, 2014
Date of Event
August 25, 2014
Report Date
August 25, 2014
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE HOMECHOICE (HC) DEVICE WAS RETURNED AND EVALUATED. THE HC UNDERWENT VISUAL EVALUATION AND ONE HOUR OF SIMULATED THERAPY WITH NO PROBLEMS FOUND. THE MACHINE HAD FULL FUNCTIONAL TESTING AND ELECTRICAL SAFETY TESTING WITH NO DEFECTS OR MALFUNCTIONS FOUND RELATED TO THE REPORTED ISSUE OF FLUID BEING OVER TEMPERATURE. THE OVER TEMPERATURE ISSUE WAS NOT VERIFIED THROUGH THE SAMPLE EVALUATION. THE EVENT HISTORY LOG REVIEW SHOWED NO KEYSTROKES, PROGRAMMING, OR USE RELATED EVENTS THAT INDICATED AND/OR CONTRIBUTED TO THE REPORTED EVENT. A REVIEW OF THE SERVICE HISTORY REVEALED NO FAILURES OR PROBLEMS THAT WERE THE SAME AS, OR SIMILAR TO THE REPORTED CONDITION AND THERE WAS NO INDICATION THAT THE PARTS REPLACED DURING SERVICING CAUSED OR CONTRIBUTED TO THE REPORTED EVENT. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE FLUID WAS OVER HEATED ON THE HOMECHOICE PRO DEVICE. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
581682 HOMECHOICE PRO SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1