DIMENSION® CLINICAL CHEMISTRY SYSTEM
Report
- Report Number
- 2517506-2014-00341
- Event Type
- Malfunction
- Date Received
- September 19, 2014
- Date of Event
- September 15, 2014
- Report Date
- September 15, 2014
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
- Product Code
- DIP
- PMA / PMN Number
- K911056
- Removal / Correction Number
- 2517506-09/24/2014-005-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
SIEMENS HEALTHCARE DIAGNOSTICS HAS CONFIRMED CUSTOMER COMPLAINTS OF IMPRECISION AND INACCURACY WITH PTN FLEX LOT FA5154. INTERNAL TESTING HAS CONFIRMED A LOW ABSORBANCE RANGE BETWEEN THE LEVELS 1 AND 5 CALIBRATORS WITH THIS LOT CAN RESULTING IN THE POTENTIAL FOR A POSITIVE OR NEGATIVE BIAS AND IMPRECISION ACROSS THE PTN ASSAY RANGE. INTERNAL STUDIES DEMONSTRATED THE BIAS IS VARIABLE ACROSS THE ASSAY RANGE FROM 9-38%. THE HIGHEST DEGREE OF BIAS IS OBSERVED BELOW THE THERAPEUTIC RANGE OF 10-20 UG/ML (39.6-79.2 UMOL/L). THE REAGENT BLANK ABSORBANCE FOR PTN LOT FA5154 MAY ALSO GENERATE ABNORMAL REACTION FLAGS. SIEMENS ISSUED AN URGENT MEDICAL DEVICE RECALL COMMUNICATION (B)(4) DATED (B)(6) 2014 TO CUSTOMERS WHO HAD ORDERED PTN LOT FA5154 ADVISING THEM TO DISCONTINUE THE USE OF THE LOT. SIEMENS OFFERED A NO CHARGE REPLACEMENT WITH A NON-IMPACTED LOT OF PHENYTOIN FLEX REAGENT CARTRIDGES.
ORIGINAL MDR WAS SUBMITTED 09/19/2014. AS A RESULT OF INVESTIGATIONS, SIEMENS HEALTHCARE DIAGNOSTICS HAS CONFIRMED CUSTOMER COMPLAINTS OF IMPRECISION AND INACCURACY WITH PTN FLEX LOT FA5154. INTERNAL TESTING HAS CONFIRMED A LOW ABSORBANCE RANGE BETWEEN THE LEVELS 1 AND 5 CALIBRATORS WITH THIS LOT CAN RESULTING IN THE POTENTIAL FOR A POSITIVE OR NEGATIVE BIAS AND IMPRECISION ACROSS THE PTN ASSAY RANGE. INTERNAL STUDIES DEMONSTRATED THE BIAS IS VARIABLE ACROSS THE ASSAY RANGE FROM 9-38%. THE HIGHEST DEGREE OF BIAS IS OBSERVED BELOW THE THERAPEUTIC RANGE OF 10-20 UG/ML (39.6-79.2 UMOL/L). THE REAGENT BLANK ABSORBANCE FOR PTN LOT FA5154 MAY ALSO GENERATE ABNORMAL REACTION FLAGS. SIEMENS ISSUED AN URGENT MEDICAL DEVICE RECALL COMMUNICATION 14-63 DATED SEPTEMBER 2014 TO CUSTOMERS WHO HAD ORDERED PTN LOT FA5154 ADVISING THEM TO DISCONTINUE THE USE OF THE LOT. SIEMENS OFFERED A NO CHARGE REPLACEMENT WITH A NON-IMPACTED LOT OF PHENYTOIN FLEX REAGENT CARTRIDGES.
DISCREPANT PHENYTOIN (PTN) RESULTS WERE OBTAINED DURING AN ATTEMPTED CALIBRATION OF LOT FA5154. ALL RESULTS WERE FLAGGED AS ABNORMAL REACTION ERRORS. THE CALIBRATION WAS NOT ACCEPTED AND PATIENT RESULTS WERE NOT REPORTED TO PHYSICIANS. THERE IS NO INDICATION THAT PATIENT TREATMENT WAS ALTERED OR PRESCRIBED ON THE BASIS OF DISCREPANT PTN RESULTS AND FAILURE TO CALIBRATE. THERE IS NO INDICATION OF ADVERSE IMPACT TO PATIENTS DUE TO THE FAILURE TO CALIBRATE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 581680 | DIMENSION® CLINICAL CHEMISTRY SYSTEM | PHENYTOIN FLEX® REAGENT CARTRIDGE | DIP | SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |