ANIMAS VIBE
Report
- Report Number
- 2531779-2014-26814
- Event Type
- Malfunction
- Date Received
- September 19, 2014
- Report Date
- September 8, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(4).
FOLLOW-UP # 1 DATE OF SUBMISSION 12/02/2014-DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 11/12/2014 WITH THE FOLLOWING FINDINGS:ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. A REVIEW OF THE PUMP BLACK BOX DATA REVEALED A CS 162 LOSS OF PRIME WARNING ASSOCIATED WITH A LOW, NON-ZERO FORCE. THE PUMP WAS EXERCISED FOR 24 HOURS AND THE REWIND, LOAD, AND PRIME STEPS WERE COMPLETED SUCCESSFULLY WITH NO DUPLICATED WARNINGS. THE FORCE SENSOR CALIBRATION TESTED LOW AND NOT WITHIN SPECIFICATIONS. THE PUMP CASE WAS REMOVED, AND THE FORCE SENSOR RESISTANCE WAS WITHIN SPECIFICATIONS. NO DAMAGE WAS FOUND TO THE FORCE SENSOR CIRCUIT. A CARTRIDGE CAP WAS NOT RETURNED WITH THE PUMP; A TEST CAP WAS USED TO COMPLETE INVESTIGATION.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A PRIME (LOSS OF PRIME) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 585115 | ANIMAS VIBE | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |