HOMECHOICE
Report
- Report Number
- 1416980-2014-32175
- Event Type
- Malfunction
- Date Received
- September 19, 2014
- Date of Event
- July 13, 2014
- Report Date
- August 26, 2014
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4).THE DEVICE WAS RECEIVED AND THE EVALUATION IS COMPLETE. THIS IS AN ANCILLARY SERVICE EVENT. THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED THROUGH AN EVENT HISTORY LOG REVIEW. UPON CONCLUSION OF THE INVESTIGATION, THE CAUSE WAS DETERMINED TO BEONE OR MORE CYCLES ADVANCES TO NEXT FILL WHEN SLOW / NO FLOW OCCURRED ABOVE THE MINIMUM DRAIN VOLUME THRESHOLD. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, AN INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2014 AT 23:17:54. DURING NIGHT DRAIN CYCLE THREE, THE PATIENT'S ULTRAFILTRATION READING WAS 761ML, INDICATING THE HOME PATIENT (HP) DRAINED 761ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 1200ML. NO ADDITIONAL INFORMATION IS AVAILABLE.