FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 4101017 · Received September 19, 2014

Report

Report Number
2531779-2014-26789
Event Type
Malfunction
Date Received
September 19, 2014
Report Date
September 8, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 10/28/2014 WITH THE FOLLOWING FINDINGS: VISUAL INSPECTION OF THE PUMP FOUND NO COSMETIC DAMAGES. REVIEW OF BLACK BOX DATA SHOWED THAT THE LAST BOLUS AND LAST BASAL WERE DELIVERED TWO AND THREE DAYS AFTER THE COMPLAINT DATE RESPECTIVELY. THE TOTAL DAILY DELIVERIES ADDED UP CORRECTLY AND REFLECTED THE USER¿S PROGRAMMED BASAL SETTINGS. THE PUMP POWERED UP AND FUNCTIONED PROPERLY. A REWIND/LOAD/PRIME SEQUENCE AND A 24-HOUR PUMP EXERCISE WERE EXECUTED WITHOUT INCIDENCES. THE PUMP DELIVERY ACCURACY WAS WITHIN SPECIFICATIONS. BOLUS EXERCISES WERE COMPLETED AND RECORDED CORRECTLY. THE INVESTIGATION WAS UNABLE TO VERIFY OR DUPLICATE THE REPORTED INACCURATE DELIVERY ISSUE.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A HISTORY/SETTINGS (INACCURATE DELIVERY) ISSUE. REVIEW OF BOLUS HISTORY SHOWED THAT BOLUS TOTALS ADD UP CORRECTLY AND ALL BOLUSES ARE RECORDED AS PROGRAMMED. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
584372 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 82 YR