FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 4100925 · Received September 19, 2014

Report

Report Number
2531779-2014-26742
Event Type
Malfunction
Date Received
September 19, 2014
Report Date
September 6, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE PUMP¿S BLACK BOX HISTORY SHOWED NO EVIDENCE OF A REWIND ISSUE OCCURRING. THE RETURNED BATTERY CAP WAS UNDAMAGED AND WAS USED TO COMPLETE ALL TESTING. DURING TESTING, THE PUMP PERFORMED THE REWIND, LOAD, AND PRIME STEPS WITH NO REWIND ISSUES OCCURRING. THE PUMP CASE WAS REMOVED AND NO INTERMITTENT CONDITION WAS FOUND TO THE MOTOR CIRCUIT. THE COMPLAINT THAT THE PUMP WOULD NOT STOP REWINDING WAS NOT ABLE TO BE DUPLICATED DURING INVESTIGATION. NO DEFECTS WERE FOUND.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A MOTOR (REWIND ISSUE) ISSUE. THE REPORTER ALLEGED THAT THE PUMP WOULD NOT STOP REWINDING. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
584908 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 64 YR