FDA Adverse Event Malfunction Summary report: N

ANIMAS VIBE

MDR report key: 4100716 · Received September 19, 2014

Report

Report Number
2531779-2014-26770
Event Type
Malfunction
Date Received
September 19, 2014
Report Date
September 5, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 12/16/2014 WITH THE FOLLOWING FINDINGS: DURING TESTING, THE PUMP WOULD NOT POWER ON DUE TO MOISTURE IN THE BATTERY COMPARTMENT. THE COMPLAINT THAT THE DISPLAY WAS DIM WAS NOT ABLE TO BE ADEQUATELY INVESTIGATED DUE TO THE PUMP NOT BEING ABLE TO POWER ON. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (DIM/FADING/COLORSPCTRM W/MOIST) ISSUE. THE REPORTER ALLEGED THAT THE DISPLAY WAS DIM. IT WAS NOTED THAT THERE WAS MOISTURE/CORROSION IN THE BATTERY COMPARTMENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
581833 ANIMAS VIBE INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 21 YR