FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 4100377 · Received September 18, 2014

Report

Report Number
2531779-2014-26707
Event Type
Malfunction
Date Received
September 18, 2014
Report Date
September 5, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 09/05/2014 WITH THE FOLLOWING FINDINGS: EVALUATION REVEALED THAT THE PUMP POWERS WITH RETURNED BATTERY CAP. THE KEYPAD IS UNRESPONSIVE TO USER INPUTS AT POWER UP. THE AUDIO BOLUS BUTTON COVER MISSING. INVESTIGATION SHOWS INTERNAL MOISTURE CONTAMINATION INSIDE AUDIO BOLUS BUTTON SWITCH CONTACT CAUSING BOLUS BUTTON TO BE SHORTED. THE KEYPAD WAS UNRESPONSIVE DUE TO SHORTED BOLUS BUTTON. REMOVED PUMP CASE AND THERE WAS NO EVIDENCE OF ADDITIONAL MOISTURE CONTAMINATION INSIDE PUMP. (B)(6).

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED CONTAMINATION UNDER KEYPAD. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 09/05/2014.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
579719 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 13 YR