FDA Adverse Event
Malfunction
Summary report: N
ONETOUCHPING GLUCOSEMGMTSYSTEM
MDR report key: 4100377
·
Received September 18, 2014
Report
- Report Number
- 2531779-2014-26707
- Event Type
- Malfunction
- Date Received
- September 18, 2014
- Report Date
- September 5, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 09/05/2014 WITH THE FOLLOWING FINDINGS: EVALUATION REVEALED THAT THE PUMP POWERS WITH RETURNED BATTERY CAP. THE KEYPAD IS UNRESPONSIVE TO USER INPUTS AT POWER UP. THE AUDIO BOLUS BUTTON COVER MISSING. INVESTIGATION SHOWS INTERNAL MOISTURE CONTAMINATION INSIDE AUDIO BOLUS BUTTON SWITCH CONTACT CAUSING BOLUS BUTTON TO BE SHORTED. THE KEYPAD WAS UNRESPONSIVE DUE TO SHORTED BOLUS BUTTON. REMOVED PUMP CASE AND THERE WAS NO EVIDENCE OF ADDITIONAL MOISTURE CONTAMINATION INSIDE PUMP. (B)(6).
Description of Event or Problem · 1
THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED CONTAMINATION UNDER KEYPAD. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 09/05/2014.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 579719 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 13 YR |