FDA Adverse Event Malfunction Summary report: N

ANIMAS VIBE

MDR report key: 4100362 · Received September 18, 2014

Report

Report Number
2531779-2014-26726
Event Type
Malfunction
Date Received
September 18, 2014
Report Date
September 5, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 12/12/2014 WITH THE FOLLOWING FINDINGS: DURING INVESTIGATION OF THE RETURNED PUMP, ¿CS69¿ ALARM WAS REPRODUCED BY APPLYING LIGHT PRESSURE ON (U39) CHIP. THE ¿CS69¿ FAILURE IS DEFINED AS LANGUAGE FILE CORRUPTION WHILE RETRIEVING THE LANGUAGE TEXT. THE ¿CS69¿ FAILURE WAS WRITTEN AS ¿CS87¿ FAILURE IN BLACK BOX. THE ERROR IS RESIDENT TO FLASH CHIP (U39). NO BATTERY CAP WAS RETURNED, TEST BATTERY CAP AND RETURNED CARTRIDGE CAP WERE USED TO COMPLETE THE INVESTIGATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A CALL SERVICE ALARM (CALL SERVICE ALARM ISSUE) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
579716 ANIMAS VIBE INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 15 YR