ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-26723
- Event Type
- Malfunction
- Date Received
- September 18, 2014
- Report Date
- September 4, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
FOLLOW-UP #1 DATE OF SUBMISSION 01/21/2015-PRODUCT ANALYSIS:THE DEVICE WAS RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 01/07/2014 WITH THE FOLLOWING FINDINGS:THE COMPLAINT WAS UNABLE TO BE DUPLICATED OR CONFIRMED WITH INVESTIGATION. REVIEW OF THE PUMP¿S BLACK BOX REVEALED EVIDENCE OF PARTIALLY DISCHARGED BATTERY INSTALLATION. A BATTERY COMPARTMENT CRACK WAS OBSERVED. THE RETURNED BATTERY CAP THREADS WERE FOUND TO BE DAMAGE; THE CAP WAS ABLE TO SECURE PROPERLY TO THE PUMP. THE PUMP¿S ELECTRIC POWER DRAWS WERE FOUND TO BE WITHIN SPECIFICATION. THERE WAS NO EVIDENCE OF DAMAGE OR DEFECT TO THE PUMP¿S INTERNAL COMPONENTS.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (BATTERY LIFE) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 580653 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 42 YR |