5MM MONOPOLAR HANDLE 33CM
Report
- Report Number
- 0002936485-2014-00736
- Event Type
- Malfunction
- Date Received
- September 18, 2014
- Date of Event
- August 25, 2014
- Report Date
- August 25, 2014
- Manufacturer
- STRYKER ENDOSCOPY-SAN JOSE
- Product Code
- KOG
- PMA / PMN Number
- K973259
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NH, US
- Reporter Occupation
- OTHER
Narratives
ADDITIONAL INFORMATION WILL BE PROVIDED ONCE THE INVESTIGATION HAS BEEN COMPLETED.
THE PRODUCT WAS RETURNED AND THE FAILURE MODE WAS CONFIRMED. THE FOLLOWING REPAIR AND SERVICE DIAGNOSTIC CODES WERE IDENTIFIED: SCRATCHES ON INSULATION. REPLACED HANDLE. THE FOLLOWING WAS OBSERVED: SCRATCHES ON INSULATION. THE HANDLE WAS REPLACED. THIS DOES CONFIRM THE ALLEGED FAILURE MODE OF INSULATION DAMAGE. PROBABLE ROOT CAUSE: POOR AUTOCLAVE RELIABILITY. INCORRECT STERILIZATION/REPROCESSING PROCEDURE. HANDLING PROCEDURES. CONTACT FORCES. PRODUCT USED BEYOND DEFINED USEFUL LIFE. IN SUM, THE PRODUCT WAS RETURNED AND THE FAILURE MODE WAS CONFIRMED.
IT WAS REPORTED THAT THE INSULATION HAS BEEN COMPROMISED.
IT WAS REPORTED THAT THE INSULATION HAS BEEN COMPROMISED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 580451 | 5MM MONOPOLAR HANDLE 33CM | ENDOSCOPE AND/OR ACCESSORIES | KOG | STRYKER ENDOSCOPY-SAN JOSE | 0643140D |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |