FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 4100099 · Received September 18, 2014

Report

Report Number
2531779-2014-26696
Event Type
Malfunction
Date Received
September 18, 2014
Report Date
September 8, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 10/29/2014 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE PUMP BLACK BOX DATA REVEALED LOSSES OF PRIME ASSOCIATED WITH LOW NON-ZERO FORCES AND ZERO FORCES. A LOSS OF PRIME WAS NOT DUPLICATED DURING TESTING. THE FORCE SENSOR CALIBRATION READING WAS WITHIN SPECIFICATIONS. THE PUMP CASE WAS REMOVED AND THE TRACE ON THE FORCE SENSOR FLEX WAS CRACKED AT THE PIN. NO EVIDENCE OF MOISTURE INGRESS WAS FOUND. UNRELATED TO THE COMPLAINT, THE BATTERY COMPARTMENT WAS OBSERVED TO BE CRACKED.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A PRIME (LOSS OF PRIME) ISSUE. THE REPORTER STATED THAT A LOSS OF PRIME EVENT HAD OCCURRED 3 OR MORE TIMES WITH AT LEAST 3 SEPARATE CARTRIDGES FROM 2 SEPARATE BOXES IN A 30 DAY PERIOD. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
578411 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 56 YR