EXETER V40 STEM 35.5MM
Report
- Report Number
- 0002249697-2014-03548
- Event Type
- Injury
- Date Received
- September 18, 2014
- Date of Event
- August 20, 2014
- Report Date
- September 2, 2014
- Manufacturer
- STRYKER ORTHOPAEDICS-MAHWAH
- Product Code
- JDI
- PMA / PMN Number
- K011623
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- PHYSICIAN
Narratives
AN EVENT REGARDING A FRACTURED STEM INVOLVING AN EXETER STEM WAS REPORTED. THE EVENT WAS CONFIRMED. METHOD & RESULTS: -DEVICE EVALUATION AND RESULTS: THE EXETER STEM FRACTURED IN FATIGUE WITH THE APPROXIMATE ORIGIN ON THE LATERAL SIDE OF THE STEM. THE STEM MATERIAL WAS CONSISTENT WITH THE ASTM F1586 AND ISO 5832-9 STAINLESS STEEL ALLOY. NO MATERIAL OR MANUFACTURING DEFECTS WERE OBSERVED ON THE PART FEATURES EXAMINED. -MEDICAL RECORDS RECEIVED AND EVALUATION: A REVIEW CONCLUDED ¿CYCLIC LOADING AT THE JUNCTION OF THE PROXIMAL LOOSE AND WELL-FIXED DISTAL STEM SEGMENTS RESULTED IN THE INEVITABLE FRACTURE. THERE IS NO EVIDENCE THAT MANUFACTURING OR MATERIALS DEFECTS WERE RESPONSIBLE FOR THIS STEM FRACTURE.¿ -DEVICE HISTORY REVIEW: THE REPORTED DEVICE WAS MANUFACTURED AND ACCEPTED INTO FINAL STOCK WITH NO DISCREPANCIES. -COMPLAINT HISTORY REVIEW: THERE HAVE BEEN NO OTHER EVENTS FOR THE REPORTED LOT ID. CONCLUSIONS: A MATERIAL ANALYSIS CONCLUDED THE STEM FRACTURED IN FATIGUE. NO MATERIAL OR MANUFACTURING DEFECTS WERE OBSERVED ON THE PARTS EXAMINED. ADDITIONALLY, A MEDICAL REVIEW CONCLUDED THAT CYCLIC LOADING AT THE JUNCTION OF A PROXIMALLY LOOSE AND DISTALLY WELL FIXED STEM LED TO AN INEVITABLE FRACTURE. THERE IS NO EVIDENCE TO SUGGEST THE EVENT IS DEVICE RELATED.
WHEN COMPLETED, THE EVALUATION SUMMARY WILL BE SUBMITTED IN A SUPPLEMENTAL REPORT.
IT IS REPORTED BY SURGEON OF THE HOSPITAL, THAT THE PATIENT HAVE HAD A HETP IMPLANTATION (B)(6) 2009. AFTER HE FELL DOWN ON HIS RIGHT SIDE HE CAME IN WITH PAIN (B)(6) 2014. NO FRACTURE WAS VISIBLE ON X-RAY (B)(6) 2014. AFTER SUSTAINED PAIN, WHICH EMANATED INTO THE THIGH, THEY MADE ANOTHER X-RAY TO CONTROL (B)(6) 2014. THIS TIME A FRACTURE WAS VISIBLE. THE EXCETER V40 STEM WAS BROKEN. THE PATIENT NEEDED TO UNDERGO A REVISION SURGERY (B)(6) 2014.
IT IS REPORTED BY SURGEON OF THE HOSPITAL, THAT THE PATIENT HAVE HAD A HETP IMPLANTATION (B)(6) 2009. AFTER HE FELL DOWN ON HIS RIGHT SIDE HE CAME IN WITH PAIN (B)(6) 2014. NO FRACTURE WAS VISIBLE ON X-RAY (B)(6) 2014. AFTER SUSTAINED PAIN, WHICH EMANATED INTO THE THIGH, THEY MADE ANOTHER X-RAY TO CONTROL (B)(6) 2014. THIS TIME A FRACTURE WAS VISIBLE. THE EXCETER V40 STEM WAS BROKEN. THE PATIENT NEEDED TO UNDERGO A REVISION SURGERY (B)(6) 2014.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 578333 | EXETER V40 STEM 35.5MM | IMPLANT | JDI | STRYKER ORTHOPAEDICS-MAHWAH | G2583259 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |