FDA Adverse Event Injury Summary report: N

EXETER V40 STEM 35.5MM

MDR report key: 4100053 · Received September 18, 2014

Report

Report Number
0002249697-2014-03548
Event Type
Injury
Date Received
September 18, 2014
Date of Event
August 20, 2014
Report Date
September 2, 2014
Manufacturer
STRYKER ORTHOPAEDICS-MAHWAH
Product Code
JDI
PMA / PMN Number
K011623
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AN EVENT REGARDING A FRACTURED STEM INVOLVING AN EXETER STEM WAS REPORTED. THE EVENT WAS CONFIRMED. METHOD & RESULTS: -DEVICE EVALUATION AND RESULTS: THE EXETER STEM FRACTURED IN FATIGUE WITH THE APPROXIMATE ORIGIN ON THE LATERAL SIDE OF THE STEM. THE STEM MATERIAL WAS CONSISTENT WITH THE ASTM F1586 AND ISO 5832-9 STAINLESS STEEL ALLOY. NO MATERIAL OR MANUFACTURING DEFECTS WERE OBSERVED ON THE PART FEATURES EXAMINED. -MEDICAL RECORDS RECEIVED AND EVALUATION: A REVIEW CONCLUDED ¿CYCLIC LOADING AT THE JUNCTION OF THE PROXIMAL LOOSE AND WELL-FIXED DISTAL STEM SEGMENTS RESULTED IN THE INEVITABLE FRACTURE. THERE IS NO EVIDENCE THAT MANUFACTURING OR MATERIALS DEFECTS WERE RESPONSIBLE FOR THIS STEM FRACTURE.¿ -DEVICE HISTORY REVIEW: THE REPORTED DEVICE WAS MANUFACTURED AND ACCEPTED INTO FINAL STOCK WITH NO DISCREPANCIES. -COMPLAINT HISTORY REVIEW: THERE HAVE BEEN NO OTHER EVENTS FOR THE REPORTED LOT ID. CONCLUSIONS: A MATERIAL ANALYSIS CONCLUDED THE STEM FRACTURED IN FATIGUE. NO MATERIAL OR MANUFACTURING DEFECTS WERE OBSERVED ON THE PARTS EXAMINED. ADDITIONALLY, A MEDICAL REVIEW CONCLUDED THAT CYCLIC LOADING AT THE JUNCTION OF A PROXIMALLY LOOSE AND DISTALLY WELL FIXED STEM LED TO AN INEVITABLE FRACTURE. THERE IS NO EVIDENCE TO SUGGEST THE EVENT IS DEVICE RELATED.

Additional Manufacturer Narrative · 1

WHEN COMPLETED, THE EVALUATION SUMMARY WILL BE SUBMITTED IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

IT IS REPORTED BY SURGEON OF THE HOSPITAL, THAT THE PATIENT HAVE HAD A HETP IMPLANTATION (B)(6) 2009. AFTER HE FELL DOWN ON HIS RIGHT SIDE HE CAME IN WITH PAIN (B)(6) 2014. NO FRACTURE WAS VISIBLE ON X-RAY (B)(6) 2014. AFTER SUSTAINED PAIN, WHICH EMANATED INTO THE THIGH, THEY MADE ANOTHER X-RAY TO CONTROL (B)(6) 2014. THIS TIME A FRACTURE WAS VISIBLE. THE EXCETER V40 STEM WAS BROKEN. THE PATIENT NEEDED TO UNDERGO A REVISION SURGERY (B)(6) 2014.

Description of Event or Problem · 1

IT IS REPORTED BY SURGEON OF THE HOSPITAL, THAT THE PATIENT HAVE HAD A HETP IMPLANTATION (B)(6) 2009. AFTER HE FELL DOWN ON HIS RIGHT SIDE HE CAME IN WITH PAIN (B)(6) 2014. NO FRACTURE WAS VISIBLE ON X-RAY (B)(6) 2014. AFTER SUSTAINED PAIN, WHICH EMANATED INTO THE THIGH, THEY MADE ANOTHER X-RAY TO CONTROL (B)(6) 2014. THIS TIME A FRACTURE WAS VISIBLE. THE EXCETER V40 STEM WAS BROKEN. THE PATIENT NEEDED TO UNDERGO A REVISION SURGERY (B)(6) 2014.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
578333 EXETER V40 STEM 35.5MM IMPLANT JDI STRYKER ORTHOPAEDICS-MAHWAH G2583259

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention