6.5CM ADULT CRANI ATTACHMT
Report
- Report Number
- 1045834-2014-12838
- Event Type
- Malfunction
- Date Received
- September 16, 2014
- Date of Event
- September 5, 2014
- Report Date
- September 5, 2014
- Manufacturer
- DEPUY SYNTHES POWER TOOLS
- Product Code
- HBC
- PMA / PMN Number
- PK011444
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- SERVICE AND TESTING PERSONNEL
Narratives
THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE AND FOUND THE DEVICE TO HAD TEMPERATURE ABOVE SPECIFICATION. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO NORMAL WEAR AND SERVICING OVER TIME. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
IT WAS REPORTED THAT DURING ENGINEERING EVALUATION, IT WAS OBSERVED THAT THE CRANIOTOME DEVICE HAD TEMPERATURE ABOVE SPECIFICATION. THIS EVENT WAS NOT RELATED TO SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO INJURIES OR MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION ASSOCIATED WITH THIS EVENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 571334 | 6.5CM ADULT CRANI ATTACHMT | MOTOR, DRILL, ELECTRIC - CRANIOTOME | HBC | DEPUY SYNTHES POWER TOOLS | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |