FDA Adverse Event
Malfunction
Summary report: N
G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
MDR report key: 4093916
·
Received September 16, 2014
Report
- Report Number
- 3004753838-2014-29227
- Event Type
- Malfunction
- Date Received
- September 16, 2014
- Date of Event
- October 18, 2013
- Report Date
- October 18, 2013
- Manufacturer
- DEXCOM, INC.
- Product Code
- MDS
- PMA / PMN Number
- P120005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT FAMILY MEMBER OR FRIEND
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE RETURNED COMPLAINT DEVICE WAS VISUALLY INSPECTED AND CUSTOMER COMPLAINT WAS CONFIRMED. THIS COMPLAINT WAS DEEMED REPORTABLE UPON COMPLETION OF DEVICE EVALUATION ON (B)(4) 2014. THE ROOT CAUSE WAS DETERMINED TO BE A DEFECTIVE COMPONENT IN THE RECEIVER'S PRINTED CIRCUIT BOARD.
Description of Event or Problem · 1
PATIENT'S FATHER CONTACTED DEXCOM TECHNICAL SUPPORT ON (B)(6) 2013 TO REPORT A CALL TECH SUPPORT ERROR MESSAGE PATIENT EXPERIENCED ON (B)(6) 2013. PATIENT'S FATHER DID NOT REPORT ANY INJURY OR MEDICAL INTERVENTION AT THE TIME OF CONTACT WITH DEXCOM TECHNICAL SUPPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 571514 | G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM | MDS | MDS | DEXCOM, INC. | MT20649 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 3 YR |