FDA Adverse Event Malfunction Summary report: N

G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM

MDR report key: 4093912 · Received September 16, 2014

Report

Report Number
3004753838-2014-29228
Event Type
Malfunction
Date Received
September 16, 2014
Date of Event
November 3, 2013
Report Date
November 3, 2013
Manufacturer
DEXCOM, INC.
Product Code
MDS
PMA / PMN Number
P120005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE RETURNED COMPLAINT DEVICE WAS VISUALLY INSPECTED AND CUSTOMER COMPLAINT WAS CONFIRMED. THIS COMPLAINT WAS DEEMED REPORTABLE UPON COMPLETION OF DEVICE EVALUATION ON (B)(6) 2014. THE ROOT CAUSE WAS DETERMINED TO BE A DEFECTIVE COMPONENT IN THE RECEIVER'S PRINTED CIRCUIT BOARD.

Description of Event or Problem · 1

PATIENT CONTACTED DEXCOM TECHNICAL SUPPORT ON (B)(6) 2013 TO REPORT A CALL TECH SUPPORT ERROR MESSAGE ON (B)(6) 2013. PATIENT DID NOT REPORT ANY INJURY OR MEDICAL INTERVENTION AT THE TIME OF CONTACT WITH DEXCOM TECHNICAL SUPPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
571329 G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM MDS MDS DEXCOM, INC. MT20649

Patients

Seq Age Sex Outcome Treatment
1 73 YR