FDA Adverse Event Malfunction Summary report: N

ENSEAL G2 STRAIGHT JAW

MDR report key: 4093798 · Received September 16, 2014

Report

Report Number
3005075853-2014-06399
Event Type
Malfunction
Date Received
September 16, 2014
Date of Event
August 28, 2014
Report Date
August 29, 2014
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GEI
PMA / PMN Number
K112033
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL INFORMATION: THE DEVICE WAS RECEIVED IN GOOD PHYSICAL CONDITION. THE DEVICE WAS TESTED ON THE GENERATOR AND PASSED ALL FUNCTIONAL TESTING. THE ENERGY OUTPUT DELIVERED FROM THE DEVICE WAS VERIFIED. ALL THREE TONES WERE HEARD DURING FUNCTIONAL TESTING (TONE 1 IS HEARD WHEN THE ENERGY ACTIVATION BUTTON IS PRESSED; TONE 2 IS HEARD WHEN TISSUE IMPEDANCE THRESHOLD IS REACHED AND TONE 3 IS HEARD WHEN THE CYCLE IS COMPLETE). NO ISSUES WERE NOTED WITH OPENING AND CLOSING OF THE JAWS. THERE WERE NO ANOMALIES NOTED WITH THE FUNCTIONALITY OF THE DEVICE. THIS REPORT IS NOT INTENDED TO DENY THAT YOU EXPERIENCED A PROBLEM WITH THE DEVICE. THERE MAY HAVE BEEN OTHER CIRCUMSTANCES OR ISSUES THAT OCCURRED DURING THE USE OF THE DEVICE THAT WE WERE UNABLE TO DUPLICATE DURING OUR LABORATORY ANALYSIS.

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME. NO DEVICE RECEIVED FOR ANALYSIS AT TIME OF SUBMISSION OF 3500A. DEVICE EVALUATION - NO DEVICE RECEIVED FOR ANALYSIS AT TIME OF SUBMISSION OF 3500A. WHEN ADDITIONAL INFORMATION IS RECEIVED AND/OR THE DEVICE ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL MEDWATCH WILL BE SENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC HYSTERECTOMY PROCEDURE, THE JAWS WERE STUCK WHILE SEALING A TISSUE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
572835 ENSEAL G2 STRAIGHT JAW ELECTROSURGICAL CUTTING & COAGULATION & ACCESSORIES GEI ETHICON ENDO-SURGERY, LLC. NA L4ET5D

Patients

Seq Age Sex Outcome Treatment
1 GENERATOR