FDA Adverse Event Malfunction Summary report: N

XMAX MOTOR

MDR report key: 4093796 · Received September 16, 2014

Report

Report Number
1045834-2014-12872
Event Type
Malfunction
Date Received
September 16, 2014
Date of Event
September 5, 2014
Report Date
September 5, 2014
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
ERL
PMA / PMN Number
PK965080
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
SERVICE AND TESTING PERSONNEL

Narratives

Additional Manufacturer Narrative · 1

THIS DEVICE WAS RETURNED FOR SERVICE HOWEVER DID NOT MEET MANUFACTURING SPECIFICATIONS DURING PRE-REPAIR ASSESSMENT. RELIABILITY ENGINEERING EVALUATED THE DEVICE AND THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO NORMAL USE AND SERVICING OVER TIME. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING SERVICE AND REPAIR PRE-TESTING, IT WAS OBSERVED THAT THE MOTOR DEVICE HAD NO ROTATION. THE EVENT WAS NOT RELATED TO SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
573245 XMAX MOTOR DRILL, SURGICAL, ENT (PNEUMATIC) - HANDPIECE ERL DEPUY SYNTHES POWER TOOLS NA

Patients

Seq Age Sex Outcome Treatment
1