FDA Adverse Event
Malfunction
Summary report: N
PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL)
MDR report key: 4093784
·
Received September 16, 2014
Report
- Report Number
- 1031452-2014-09452
- Event Type
- Malfunction
- Date Received
- September 16, 2014
- Report Date
- August 25, 2014
- Manufacturer
- INVACARE FLORIDA OPERATIONS
- Product Code
- CAW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
Description of Event or Problem · 1
PER PROVIDER THE VALVE IS STUCK. PER INDEPENDENT REPAIR CENTER THE UNIT HAS LOW O2/YELLOW LIGHT. THE REXROTH VALVE IS STUCK. THE POPPIT KIT VALVE IS NOT SHIFTING AND THE SIEVE BEDS IS SATURATED. THE TUBING ON THE UNIT IS LEAKING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 573241 | PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL) | 868.5440 | CAW | INVACARE FLORIDA OPERATIONS | IRC5PO2 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |