FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 4093744 · Received September 16, 2014

Report

Report Number
2531779-2014-26450
Event Type
Malfunction
Date Received
September 16, 2014
Report Date
September 1, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE DEVICE EVALUATION WAS COMPLETED FOR A PREVIOUS COMPLAINT ON 12/10/2013, PRIOR TO THE DATE THIS COMPLAINT WHICH WAS REPORTED TO ANIMAS, ON (B)(6) 2014. PRODUCT ANALYSIS FROM THE PREVIOUS INVESTIGATION HAD THE FOLLOWING FINDINGS: THE RETURNED BATTERY CAP WAS ABLE TO FULLY TIGHTEN TO THE PUMP AND WAS ABLE TO SUCCESSFULLY MAINTAIN ELECTRICAL CONNECTION. THE PUMP WAS ABLE TO SUCCESSFULLY COMPLETE THE REWIND, LOAD, AND PRIME STEPS. NO ADDITIONAL TESTING COULD BE COMPLETED BECAUSE THE PUMP WAS DISASSEMBLED AS PART OF THE PRIOR TESTING AND THE PUMP WAS NO LONGER AVAILABLE TO FULLY INVESTIGATE THE REPORTED MOTOR ISSUE.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A MOTOR (MOTOR ISSUE) ISSUE. THE REPORTER STATED THAT THE REWIND AND LOAD STEPS HAD BEEN TAKING MUCH LONGER THAN EXPECTED CAUSING THE PUMP TO GO TO SLEEP AND THEN BACK TO THE HOME SCREEN. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
573605 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 14 YR