FDA Adverse Event Malfunction Summary report: N

PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL)

MDR report key: 4093738 · Received September 16, 2014

Report

Report Number
1031452-2014-09446
Event Type
Malfunction
Date Received
September 16, 2014
Report Date
August 22, 2014
Manufacturer
INVACARE FLORIDA OPERATIONS
Product Code
CAW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Description of Event or Problem · 1

PER INDEPENDENT REPAIR CENTER STATEMENT, THE UNIT HAD LOW O2 OR YELLOW LIGHT. THE KEY FAILURE WAS A LOOSE CLAMP ON THE SIEVE BED/MANIFOLD THAT WAS REPLACED. ADDITIONAL MALFUNCTIONS WERE THE INLET FILTER WAS DIRTY, THE BLUE ZIP TIE ON THE SIEVE BED/MANIFOLD NEEDED TO BE REPLACED.