FDA Adverse Event Injury Summary report: N

CHARGING SYSTEM

MDR report key: 4093664 · Received September 4, 2014

Report

Report Number
1627487-2014-21591
Event Type
Injury
Date Received
September 4, 2014
Date of Event
June 16, 2014
Report Date
August 12, 2014
Manufacturer
ST. JUDE MEDICAL - NEUROMODULATION
Product Code
GZB
PMA / PMN Number
P010032
Removal / Correction Number
1627487-07262012-001-C
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THIS CHARGER MODEL WAS ASSOCIATED WITH A FIELD CORRECTION. MFR'S EVAL: CORRECTIVE AND PREVENTIVE ACTION CAPA) INVESTIGATION WAS PERFORMED. RESULT: POCKET HEATING CONFIRMED. THE INVESTIGATION FOR CAPA (B)(4) ASSOCIATED WITH HEATING WHILE CHARGING (POCKET HEATING) CONCLUDED THAT THE CHARGER WAS CAPABLE OF TRANSFERRING ENERGY TO THE IPG AT A RATE THAT WOULD CAUSE HEATING OF THE IPG AND/OR CHARGING WAND OF SUFFICIENT ELEVATED TEMPERATURE TO CAUSE PAIN AND BURNS. THE HEATING WHILE CHARGING WAS DETERMINED TO BE EXACERBATED BY OFF-AXIS CHARGING OF SHALLOW IMPLANTED IPGS AND THAT ALL CHARGERS WERE CAPABLE OF ELEVATED HEATING. SJM HAS LIMITED INFO RELATED TO THE PT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Description of Event or Problem · 1

DEVICE 2 OF 2. REFERENCE MFR. REPORT: 1627487-2014-21590. THE PT REPORTED EXPERIENCING POCKET HEATING WHILE CHARGING THE IPG. A NEW LOW ENERGY CHARGING SYSTEM WAS SENT TO THE PT. ON 8/1/2012 ST. JUDE MEDICAL, NEUROMODULATION DIVISION, SENT FIELD ACTION LETTERS TO PTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS WAS EXPECTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
541589 CHARGING SYSTEM SCS CHARGING SYSTEM GZB ST. JUDE MEDICAL - NEUROMODULATION 3701 4590163

Patients

Seq Age Sex Outcome Treatment
1 50 YR Other SCS LEAD, MODEL 3289| SCS LEAD, MODEL 3288