FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 4093662 · Received September 16, 2014

Report

Report Number
2531779-2014-26445
Event Type
Malfunction
Date Received
September 16, 2014
Report Date
September 4, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP #1; DATE OF SUBMISSION: (B)(4) 2014. DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED, AND IT WAS EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: INSPECTION REVEALED THAT THE DISPLAY SCREEN VISUAL WAS DIM AND DISCOLORED: THE REPORTED ISSUE WAS CONFIRMED. THE REPORTED ISSUE WAS DIRECTLY CAUSED BY AN UNIDENTIFIED DISPLAY FAILURE. UNRELATED TO THE REPORTED ISSUE, THE AUDIO BOLUS BUTTON (ABB) COVER WAS OBSERVED TO BE MISSING, AND THUS THE ABB WAS NOT OPERATIONAL.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (DIM/FADING/COLOR SPECTRUM) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.