FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 4093633 · Received September 16, 2014

Report

Report Number
1416980-2014-31556
Event Type
Malfunction
Date Received
September 16, 2014
Date of Event
August 17, 2014
Report Date
August 22, 2014
Manufacturer
BAXTER HEALTHCARE ? SINGAPORE
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL METHOD: THE DEVICE WAS RETURNED TO BAXTER AND THE EVALUATION IS COMPLETE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT. THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED THROUGH AN EVENT HISTORY LOG REVIEW. ALL RETURNED HOMECHOICE DEVICES RECEIVE A RETURNED INSTRUMENT TESTING EVALUATION (RITE). THIS EVALUATION INCLUDES VISUAL INSPECTION, FUNCTIONAL AND ELECTRICAL TESTING ON THE DEVICE. A SHORT SIMULATED THERAPY WAS PERFORMED SUCCESSFULLY. THE HOMECHOICE PASSED ALL RETURNED INSTRUMENT TESTING EVALUATION PERFORMED. UPON CONCLUSION OF THE INVESTIGATION, THE CAUSE WAS DETERMINED TO BE USE ERROR, TIDAL TOTAL ULTRAFILTRATION (UF) REMOVAL SET TOO LOW. HOMECHOICE APD SYSTEMS TRAINER¿S GUIDE GIVES INSTRUCTIONS ON HOW TO SET THE TIDAL THERAPY SETTINGS. A FORMAL REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2014 23:44:44. DURING NIGHT DRAIN CYCLE THREE, THE PATIENT'S ULTRAFILTRATION READING WAS 1679ML, INDICATING THE HOME PATIENT (HP) DRAINED 1349ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 2200ML. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
572908 HOMECHOICE PRO SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE ? SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1 54 YR