FDA Adverse Event Injury Summary report: N

CHARGING SYSTEM

MDR report key: 4093577 · Received September 4, 2014

Report

Report Number
1627487-2014-12579
Event Type
Injury
Date Received
September 4, 2014
Date of Event
September 1, 2012
Report Date
August 8, 2014
Manufacturer
ST. JUDE MEDICAL - NEUROMODULATION
Product Code
GZB
PMA / PMN Number
P010032
Removal / Correction Number
1627487-07262012-001-C
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THIS CHARGER MODEL NUMBER WAS INCLUDED IN A FIELD CORRECTION. CORRECTIVE AND PREVENTIVE ACTION (CAPA). RESULT: POCKET HEATING WAS CONFIRMED. THE INVESTIGATION FOR CAPA (B)(4) ASSOCIATED WITH HEATING WHILE CHARGING (POCKET HEATING) CONCLUDED THAT THE CHARGER WAS CAPABLE OF TRANSFERRING ENERGY TO THE IPG AT A RATE THAT WOULD CAUSE HEATING OF THE IPG AND/OR CHARGING WAND OF SUFFICIENT ELEVATED TEMPERATURE TO CAUSE PAIN AND BURNS. THE HEATING WHILE CHARGING WAS DETERMINED TO BE EXACERBATED BY OFF-AXIS CHARGING OF SHALLOW IMPLANTED IPGS AND THAT ALL CHARGES WERE CAPABLE OF ELEVATED HEATING. SJM HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Description of Event or Problem · 1

DEVICE 2 OF 2. REFERENCE MFR REPORT #1627487-2014-12578.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
542031 CHARGING SYSTEM SCS CHARGER GZB ST. JUDE MEDICAL - NEUROMODULATION 3701 83962

Patients

Seq Age Sex Outcome Treatment
1 45 YR Other SCS LEAD, MODEL: 3186| IMPLANT DATE: