FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 4093376 · Received September 16, 2014

Report

Report Number
1416980-2014-31514
Event Type
Malfunction
Date Received
September 16, 2014
Date of Event
June 8, 2014
Report Date
August 22, 2014
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND THE EVALUATION IS COMPLETE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT. THE HIGH DRAIN 104 ALARM WAS IDENTIFIED IN THE EVENT HISTORY LOG REVIEW. ALL RETURNED HOMECHOICE DEVICES RECEIVE A RETURNED INSTRUMENT TESTING EVALUATION (RITE). THIS EVALUATION INCLUDES VISUAL INSPECTION, FUNCTIONAL AND ELECTRICAL TESTING ON THE DEVICE. A SERVICE HISTORY REVIEW SHOWED NO FAILURES/PROBLEMS THAT WERE THE SAME AS, OR SIMILAR TO, THE CURRENT DIFFICULTY. IN ADDITION, THERE WAS NO INDICATION THAT THE PARTS REPLACED DURING SERVICING CAUSED OR CONTRIBUTED TO THE REPORTED DIFFICULTY. THE CAUSE COULD NOT BE DETERMINED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RECEIVED AND IS CURRENTLY IN THE PROCESS OF BEING EVALUATED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING A REVIEW OF THE LOGS OF A RETURNED HOMECHOICE (HC) MACHINE, AN INCREASED INTRA-PERITONEAL VOLUME (IIPV) WAS IDENTIFIED AS A HIGH DRAIN 104 (MANUAL DRAIN 4) ALARM. THIS ALARM OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2014 AT 17:45:36. THIS ALARM INDICATES THE PATIENT DRAINED GREATER THAN 200% OF THE MAXIMUM PRESCRIBED CYCLE FILL VOLUME. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
571434 HOMECHOICE PRO SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1 57 YR