ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-26401
- Event Type
- Malfunction
- Date Received
- September 16, 2014
- Report Date
- September 3, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- PATIENT
Narratives
DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 10/30/2014 WITH THE FOLLOWING FINDINGS: THERE WERE PUMP REBOOTS OBSERVED IN THE BLACK BOX HISTORY IN REFERENCE WITH CLEARING ALARMS. THERE WERE CALL SERVICE (CS 052 AND CS054) ALARMS OBSERVED IN THE BLACK BOX HISTORY. UNRELATED TO THE ORIGINAL COMPLAINT, THE PUMP CASE WAS CRACKED AND WAS CONFIRMED WITH THE LEAK TEST. THE PUMP CASE WAS REMOVED AND THERE WAS EVIDENCE OF MOISTURE CONTAMINATION THROUGHOUT THE PUMP. ALSO UNRELATED, THERE WAS MOISTURE EVIDENCE BEHIND THE DISPLAY LENS. DUE TO INTERNAL MOISTURE CONTAMINATION, THE PUMP POWERED ON WITH AUDIO AND VIBRATION, BUT HAD AN UNREADABLE MULTI-COLORED DISTORTED DISPLAY IMAGE. THE BATTERY CAP WAS ABLE TO FULLY TIGHTEN AND THE BATTERY COMPARTMENT WAS INTACT. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (NO POWER) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 571180 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 39 YR |