FDA Adverse Event Malfunction Summary report: N

PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL)

MDR report key: 4093284 · Received September 16, 2014

Report

Report Number
1031452-2014-09498
Event Type
Malfunction
Date Received
September 16, 2014
Report Date
August 25, 2014
Manufacturer
INVACARE FLORIDA OPERATIONS
Product Code
CAW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Description of Event or Problem · 1

IT WAS REPORTED BY THE PROVIDER THAT THE VALVE IS STUCK. PER INDEPENDENT REPAIR CENTER STATEMENT: THE UNIT IS ALARMING OR RED LIGHT. THE REXROTH VALVE IS STUCK, THE POPPET KIT VALVE IS NOT SHIFTING, AND THE SIEVE BEDS IS SATURATED THE MUFFLER IS LEAKING, THE WARM CLAMP AND ZIP TIES TUBING ARE LEAKING. NO REPORT OF PATIENT INJURY, NO ADDITIONAL INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
571629 PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL) 868.5440 CAW INVACARE FLORIDA OPERATIONS IRC5PO2V

Patients

Seq Age Sex Outcome Treatment
1 Other