FDA Adverse Event Malfunction Summary report: N

CS300

MDR report key: 4093223 · Received August 21, 2014

Report

Report Number
2249723-2014-01214
Event Type
Malfunction
Date Received
August 21, 2014
Date of Event
April 27, 2012
Report Date
April 27, 2012
Manufacturer
DATASCOPE COR.
Product Code
DSP
PMA / PMN Number
K063525
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND ELECTRONIC ERROR CODE 51, MOISTURE DAMAGE ON THE MOTOR CONTROLLER BOARD (PART NUMBER: 0671-00-0004) AND BROKEN IV POLE RETAINING ASSEMBLY (PART NUMBER 0105-00-0114). BOTH PARTS WERE REPLACED. THE TECHNICIAN PERFORMED CALIBRATION, FUNCTIONAL TESTS AND SAFETY CHECKS. THE IABP MET FACTORY SPECIFICATIONS AND WAS RETURNED TO THE CUSTOMER. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THERE WAS A FLUID SPILL ON THE UNIT. THE PUMP WAS REPLACED. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
504547 CS300 INTRA-AORTIC BALLOON PUMP DSP DATASCOPE COR. CS300

Patients

Seq Age Sex Outcome Treatment
1