FDA Adverse Event
Malfunction
Summary report: N
CS300
MDR report key: 4093223
·
Received August 21, 2014
Report
- Report Number
- 2249723-2014-01214
- Event Type
- Malfunction
- Date Received
- August 21, 2014
- Date of Event
- April 27, 2012
- Report Date
- April 27, 2012
- Manufacturer
- DATASCOPE COR.
- Product Code
- DSP
- PMA / PMN Number
- K063525
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
THE TECHNICIAN FOUND ELECTRONIC ERROR CODE 51, MOISTURE DAMAGE ON THE MOTOR CONTROLLER BOARD (PART NUMBER: 0671-00-0004) AND BROKEN IV POLE RETAINING ASSEMBLY (PART NUMBER 0105-00-0114). BOTH PARTS WERE REPLACED. THE TECHNICIAN PERFORMED CALIBRATION, FUNCTIONAL TESTS AND SAFETY CHECKS. THE IABP MET FACTORY SPECIFICATIONS AND WAS RETURNED TO THE CUSTOMER. (B)(4).
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THERE WAS A FLUID SPILL ON THE UNIT. THE PUMP WAS REPLACED. NO PATIENT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 504547 | CS300 | INTRA-AORTIC BALLOON PUMP | DSP | DATASCOPE COR. | CS300 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |