FDA Adverse Event
Malfunction
Summary report: N
TERUMO TCM II COOLING AND HEATING SYSTEM
MDR report key: 4093166
·
Received August 21, 2014
Report
- Report Number
- 1828100-2014-00682
- Event Type
- Malfunction
- Date Received
- August 21, 2014
- Date of Event
- July 1, 2014
- Report Date
- July 30, 2014
- Manufacturer
- TERUMO CARDIOVASCULAR SYSTEMS CORP.
- Product Code
- DWC
- PMA / PMN Number
- K883603
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
PER THE CUSTOMER, THERE WERE NO LEAKS, LIGHTS OR ANY HEATING/COOLING PROBLEMS WITH THE UNIT.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING SET-UP/PRIMING OF THE DEVICE FOR A CARDIOPULMONARY BYPASS PROCEDURE, THE COOLER HEATER UNIT'S ICE BLOCK WAS TOO LARGE. THE BLOCK WAS SO BIG THAT WATER COULD NOT CIRCULATE THROUGH THE SYSTEM AND THEY COULD NOT AUTOMATICALLY DEFROST THE BLOCK. AS A RESULT, AN ALTERNATIVE DEVICE WAS EMPLOYED. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WERE NO DELAYS, NO BLOOD LOSS, AND NO ADVERSE CONSEQUENCES TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 506878 | TERUMO TCM II COOLING AND HEATING SYSTEM | DWC | TERUMO CARDIOVASCULAR SYSTEMS CORP. | 4415 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |