FDA Adverse Event Malfunction Summary report: N

TERUMO TCM II COOLING AND HEATING SYSTEM

MDR report key: 4093166 · Received August 21, 2014

Report

Report Number
1828100-2014-00682
Event Type
Malfunction
Date Received
August 21, 2014
Date of Event
July 1, 2014
Report Date
July 30, 2014
Manufacturer
TERUMO CARDIOVASCULAR SYSTEMS CORP.
Product Code
DWC
PMA / PMN Number
K883603
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

PER THE CUSTOMER, THERE WERE NO LEAKS, LIGHTS OR ANY HEATING/COOLING PROBLEMS WITH THE UNIT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING SET-UP/PRIMING OF THE DEVICE FOR A CARDIOPULMONARY BYPASS PROCEDURE, THE COOLER HEATER UNIT'S ICE BLOCK WAS TOO LARGE. THE BLOCK WAS SO BIG THAT WATER COULD NOT CIRCULATE THROUGH THE SYSTEM AND THEY COULD NOT AUTOMATICALLY DEFROST THE BLOCK. AS A RESULT, AN ALTERNATIVE DEVICE WAS EMPLOYED. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WERE NO DELAYS, NO BLOOD LOSS, AND NO ADVERSE CONSEQUENCES TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
506878 TERUMO TCM II COOLING AND HEATING SYSTEM DWC TERUMO CARDIOVASCULAR SYSTEMS CORP. 4415

Patients

Seq Age Sex Outcome Treatment
1