ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-26385
- Event Type
- Malfunction
- Date Received
- September 16, 2014
- Report Date
- September 3, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- PATIENT
Narratives
DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED, AND IT WAS EVALUATED BY PRODUCT ANALYSIS ON 11/05/2014 WITH THE FOLLOWING FINDINGS: VISUAL INSPECTION REVEALED THAT THE KEYPAD COVER WAS PEELING FROM ITS BASE. EVALUATION REVEALED THAT THE CONTRAST KEYPAD BUTTON WAS UNRESPONSIVE: THE REPORTED ISSUE WAS NOT DUPLICATED, AS AN UNDER-RESPONSIVE CONDITION OF THE UP ARROW, DOWN ARROW, AND/OR OK KEYPAD BUTTONS WAS NOT FOUND. THE KEYPAD COVER WAS REMOVED, AND THE CONTRAST KEY CONTACT WAS OBSERVED TO BE INVERTED; THIS DEVICE FAILURE CAUSED THE OBSERVED KEYPAD-RESPONSIVENESS ISSUE. THE UP ARROW, DOWN ARROW, AND OK KEY CONTACTS WERE OBSERVED TO BE FREE OF DAMAGE, DEFECT, AND CONTAMINATION. THE OBSERVED DEVICE FAILURES ARE UNRELATED TO THE REPORTED ISSUE. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE. THE REPORTER ALLEGED THAT THE UP ARROW, DOWN ARROW, AND OK KEYPAD BUTTONS WERE UNDER-RESPONSIVE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 573074 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 13 YR |