LIFEVEST WCD 4000 SYSTEM

MDR report key: 4093067 · Received September 10, 2014

Report

Report Number
3008642652-2014-02776
Event Type
Malfunction
Date Received
September 10, 2014
Date of Event
August 15, 2014
Report Date
September 5, 2014
Manufacturer
ZOLL MANUFACTURING CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (UNABLE TO COMMUNICATE WITH MONITOR) HAS BEEN CONFIRMED. AS RECEIVED, THE BELT FAILED INCOMING TESTING. UPON EVALUATION, THERE WAS AN OPENING ALONG THE FRONT PULSE WIRE AND THE BROWN (-5V) WIRE OF THE TRUNK CABLE. THE CAUSE OF THE TEST FAILURE IS THE OPEN WIRES. THE ROOT CAUSE OF THE OPEN WIRES CANNOT BE POSITIVELY IDENTIFIED BUT IS LIKELY EXCESSIVE FORCE PLACED ON THE CABLE. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE BELT.

Description of Event or Problem · 1

A US DISTRIBUTOR CONTACTED ZOLL TO REPORT THAT ELECTRODE BELT SN (B)(4) WAS UNABLE TO COMMUNICATE WITH A MONITOR.