PORTEX ANESTHESIA BREATHING CIRCUIT
Report
- Report Number
- 2183502-2014-00648
- Event Type
- Malfunction
- Date Received
- September 8, 2014
- Date of Event
- June 17, 2014
- Report Date
- September 5, 2014
- Manufacturer
- SMITHS MEDICAL INTERNATIONAL LTD
- Product Code
- CAI
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
THREE UNUSED CORRUGATED TUBES, FROM THE SAME LOT, WERE RETURNED FOR PRODUCT EVAL. DURING VISUAL INSPECTION, NO CRACKS, HOLES, OR DEFECTS WERE DISCOVERED IN ANY OF THE RETURNED PRODUCTS. EACH TUBE WAS FUNCTIONALLY TESTED. NO LEAKS WERE OBSERVED; THE TUBES WERE FOUND TO OPERATE AS INTENDED. THE TUBES WERE THEN TENSILE TESTED. ALL STRENGTH TESTING RESULTS PASSED SPEC. THE REPORTED PRODUCT FAULT COULD NOT BE DUPLICATED. NO EVIDENCE WAS FOUND TO SUGGEST THE REPORTED EVENT WAS CAUSED BY AN INTRINSIC PRODUCT FAULT.
A REPORT WAS RECEIVED STATING THAT THE LISTED DEVICE WAS IN USE FOR 15 MINUTES WHEN THE VENTILATOR ALARMED. THE USER REPORTEDLY FOUND A TEAR IN CORRUGATED TUBING, WHICH CAUSED A LEAK IN THE SYSTEM AND THE ALARM ON THE VENTILATOR. THERE WERE NO ADVERSE EFFECTS TO THE PT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 550988 | PORTEX ANESTHESIA BREATHING CIRCUIT | CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE) | CAI | SMITHS MEDICAL INTERNATIONAL LTD | NA | 2539102 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |