OPTIQ CATHETER
Report
- Report Number
- 2025816-2014-00103
- Event Type
- Malfunction
- Date Received
- September 4, 2014
- Date of Event
- July 10, 2014
- Report Date
- July 11, 2014
- Manufacturer
- ICU MEDICAL, INC.
- Product Code
- FPA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- HR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
DEVICE RETURN: THE INVOLVED 011-52509-12 WAS NOT RETURNED FOR ANALYSIS. SIX (6) PACKAGED 011-52509-12 SAME LOT 2768898 SAMPLES WERE RETURNED. ENGINEERING INVESTIGATION, TESTING AND ANALYSIS: THE RETURNED 011-52509-12 CATHETERS WERE TESTED PER THE APPLICABLE PRODUCT SPECS INCLUDING THERMISTOR ACCURACY TEST, THERMAL COIL CIRCUIT RESISTANCE CHECK. THE RESULTS RECORDED NO OUT OF SPEC CONDITIONS AND OR PERFORMANCE ANOMALIES. LOT BUILD REVIEWS: A REVIEW OF THE MFG LOT BUILD DATABASE FOR THE 011-52509-12 LOT# 2768898 (MFG DATE 11/01/2013) SHOWS (B)(4) UNITS WERE ALL MFG., TESTED, INSPECTED AND RELEASED. THERE WERE NO EXCEPTION DOCUMENTS GENERATED DURING THE MFG. LOT BUILDS. FINDINGS: THE INVOLVED 011-52509-12 WAS NOT RETURNED FOR TESTING AND ANALYSIS. ENGINEERING TESTING OF THE SIX RETURNED SAME LOT SAMPLES RECORDED NO PERFORMANCE ISSUES AND/OR PUT OF SPEC CONDITIONS. THE EXACT CAUSE(S) OF THE REPORTED EVENT/PRODUCT ISSUES IS UNK AT THIS TIME. THIS REPORT AND THE ASSOCIATED DATA HAVE BEEN ENTERED IN THE MANUFACTURERS DATABASE AND WILL BE MONITORED AND TRENDED.
INT'L ((B)(6)) COMPLAINT RECEIVED REPORTING ERRONEOUS READINGS WITH USE OF (B)(6) OPTIQ CATHETER. IT WAS REPORTED THAT "...DURING THE OPERATION, MEASURED CCO WAS NOT REALISTIC, IT WAS TOO LOW AND SIGNAL WAS CONTINUOUSLY LOST. THE CLINICAL PICTURE OF THE PT HEMODYNAMIC PARAMETERS WERE NORMAL, SUCH AS HR, MAP (SYSTOLE, DIASTOLE), CVP, SPO2. HB, ETC." MEASUREMENTS WERE DONE BY BOLUS. FOLLOW UP INFO REPORTED "..WHEN THE CATHETER WAS CONNECTED TO THE ANOTHER MONITOR AFTER OPERATION, NORMAL VALUES WERE SHOWED.". THERE WERE NO ADVERSE PT CONSEQUENCES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 542171 | OPTIQ CATHETER | TD CATHETER | FPA | ICU MEDICAL, INC. | 011-52509-12 | 2768898 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NI | Q2 PLUS MONITOR,| CABLES |