ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-26371
- Event Type
- Malfunction
- Date Received
- September 16, 2014
- Report Date
- September 3, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UT, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 10/14/2014 WITH THE FOLLOWING FINDINGS: THE PUMP WAS RETURNED WITH A PEELING KEYPAD COVER. DURING TESTING, THE UP AND DOWN ARROW KEYPAD BUTTONS WERE INTERMITTENTLY RESPONSIVE, AND THE CONTRAST BUTTON WAS UNRESPONSIVE. THE KEYPAD COVER WAS REMOVED AND CONTAMINATION WAS FOUND UNDER ALL OF THE KEYPAD BUTTON CONTACTS. UNRELATED TO THE INITIAL COMPLAINT, THE AUDIO BOLUS BUTTON COVER WAS FOUND TO BE TORN AND EXPOSING THE WHITE SLUG; HOWEVER, THE BOLUS BUTTON WAS NORMALLY RESPONSIVE. ALSO UNRELATED, THE DISPLAY SCREEN WAS DIM AND DISCOLORED, AND THE CARTRIDGE COMPARTMENT WAS OBSERVED TO BE CRACKED. THERE WAS NO VISIBLE DAMAGE TO THE CARTRIDGE CAP AND WAS ABLE TO FULLY SECURE ONTO THE PUMP.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (DAMAGE PRIOR TO TACTILE CHANGE) ISSUE. THE REPORTER ALLEGED THAT THE UP ARROW KEYPAD BUTTON WAS OVER RESPONSIVE, AND THAT THE KEYPAD COVER WAS MISSING/PEELING AND TORN/PUNCTURED. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 571418 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 23 YR |