FDA Adverse Event Malfunction Summary report: N

TRILOGY 100

MDR report key: 4092786 · Received August 20, 2014

Report

Report Number
2518422-2014-01390
Event Type
Malfunction
Date Received
August 20, 2014
Date of Event
July 28, 2014
Report Date
July 28, 2014
Manufacturer
RESPIRONICS, INC
Product Code
CBK
PMA / PMN Number
K083526
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DURING THE EVAL FOR THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, A "SERVICE REQUIRED" CODE WAS FOUND IN THE VENTILATOR'S DOWNLOADED ERROR LOG. THE DEVICE'S POWER MANAGEMENT BOARD WILL BE REPLACED TO ADDRESS THE ISSUE. A "VENTILATOR INOPERATIVE" CODE WAS FOND IN THE VENTILATOR'S DOWNLOADED ERROR LOG. THE DEVICE'S SYSTEM BOARD WILL BE REPLACED TO ADDRESS THE ISSUE. DEVICE HAS BEEN EVALUATED BUT THE COMPONENTS HAVE NOT BEEN REPLACED PENDING CUSTOMER APPROVAL OF THE ESTIMATE.

Description of Event or Problem · 1

THE MANUFACTURER RECEIVED INFO ALLEGING A "SERVICE REQUIRED" ALARM CONDITION OCCURRED. THERE WAS NO HARM OR INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
500642 TRILOGY 100 VENTILATOR, CONTINUOUS, FACILITY USE CBK RESPIRONICS, INC 1054096

Patients

Seq Age Sex Outcome Treatment
1