FDA Adverse Event
Malfunction
Summary report: N
TRILOGY 100
MDR report key: 4092786
·
Received August 20, 2014
Report
- Report Number
- 2518422-2014-01390
- Event Type
- Malfunction
- Date Received
- August 20, 2014
- Date of Event
- July 28, 2014
- Report Date
- July 28, 2014
- Manufacturer
- RESPIRONICS, INC
- Product Code
- CBK
- PMA / PMN Number
- K083526
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DURING THE EVAL FOR THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, A "SERVICE REQUIRED" CODE WAS FOUND IN THE VENTILATOR'S DOWNLOADED ERROR LOG. THE DEVICE'S POWER MANAGEMENT BOARD WILL BE REPLACED TO ADDRESS THE ISSUE. A "VENTILATOR INOPERATIVE" CODE WAS FOND IN THE VENTILATOR'S DOWNLOADED ERROR LOG. THE DEVICE'S SYSTEM BOARD WILL BE REPLACED TO ADDRESS THE ISSUE. DEVICE HAS BEEN EVALUATED BUT THE COMPONENTS HAVE NOT BEEN REPLACED PENDING CUSTOMER APPROVAL OF THE ESTIMATE.
Description of Event or Problem · 1
THE MANUFACTURER RECEIVED INFO ALLEGING A "SERVICE REQUIRED" ALARM CONDITION OCCURRED. THERE WAS NO HARM OR INJURY REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 500642 | TRILOGY 100 | VENTILATOR, CONTINUOUS, FACILITY USE | CBK | RESPIRONICS, INC | 1054096 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |